Article ID Journal Published Year Pages File Type
4225758 European Journal of Radiology 2012 5 Pages PDF
Abstract

ObjectiveIn medical imaging, lowering radiation dose from computed tomography scanning, without reducing diagnostic performance is a desired achievement. Iterative image reconstruction may be one tool to achieve dose reduction. This study reports the diagnostic performance using a blending of 50% statistical iterative reconstruction (ASIR) and filtered back projection reconstruction (FBP) compared to standard FBP image reconstruction at different dose levels for liver phantom examinations.MethodsAn anthropomorphic liver phantom was scanned at 250, 185, 155, 140, 120 and 100 mA s, on a 64-slice GE Lightspeed VCT scanner. All scans were reconstructed with ASIR and FBP. Four readers evaluated independently on a 5-point scale 21 images, each containing 32 test sectors. In total 672 areas were assessed. ROC analysis was used to evaluate the differences.ResultsThere was a difference in AUC between the 250 mA s FBP images and the 120 and 100 mA s FBP images. ASIR reconstruction gave a significantly higher diagnostic performance compared to standard reconstruction at 100 mA s.ConclusionA blending of 50–90% ASIR and FBP may improve image quality of low dose CT examinations of the liver, and thus give a potential for reducing radiation dose.

Related Topics
Health Sciences Medicine and Dentistry Radiology and Imaging