Article ID Journal Published Year Pages File Type
4226110 European Journal of Radiology 2012 5 Pages PDF
Abstract

The purpose of this study was to compare 3 T and 7 T signal-to-noise and contrast-to noise ratios of clinical sequences for imaging of the ankles with optimized sequences and dedicated coils.Ten healthy volunteers were examined consecutively on both systems with three clinical sequences: (1) 3D gradient-echo, T1-weighted; (2) 2D fast spin-echo, PD-weighted; and (3) 2D spin-echo, T1-weighted. SNR was calculated for six regions: cartilage; bone; muscle; synovial fluid; Achilles tendon; and Kager's fat-pad. CNR was obtained for cartilage/bone, cartilage/fluid, cartilage/muscle, and muscle/fat-pad, and compared by a one-way ANOVA test for repeated measures.Mean SNR significantly increased at 7 T compared to 3 T for 3D GRE, and 2D TSE was 60.9% and 86.7%, respectively. In contrast, an average SNR decrease of almost 25% was observed in the 2D SE sequence. A CNR increase was observed in 2D TSE images, and in most 3D GRE images.There was a substantial benefit from ultra high-field MR imaging of ankles with routine clinical sequences at 7 T compared to 3 T. Higher SNR and CNR at ultra-high field MR scanners may be useful in clinical practice for ankle imaging. However, carefully optimized protocols and dedicated extremity coils are necessary to obtain optimal results.

Related Topics
Health Sciences Medicine and Dentistry Radiology and Imaging
Authors
, , , , , ,