Article ID Journal Published Year Pages File Type
4226514 European Journal of Radiology 2011 6 Pages PDF
Abstract

ObjectiveTo examine the feasibility of reducing the inter-scan variability of prospectively ECG-triggered calcium-scoring scans by using overlapping 3-mm datasets generated from multiplanar reformation (MPR) instead of non-overlapping 3-mm or 1.5-mm datasets.Patients and methodsSeventy-five women (59–79 years old) underwent two sequential prospectively ECG-triggered calcium-scoring scans with 16 mm × 1.5 mm collimation in one session. Between the two scans patients got off and on the table. We performed calcium scoring (Agatston and mass scores) on the following datasets: contiguous 3-mm sections reconstructed from the raw data (A), contiguous 3-mm sections from MPR (B), overlapping 3-mm sections from MPR (C) and contiguous 1.5-mm sections from the raw data (D). To determine the feasibility of the MPR approach, we compared MPR (B) with direct raw data reconstruction (A). Inter-scan variability was calculated for each type of dataset (A–D).ResultsCalcium scores ranged from 0 to 1455 (Agatston) and 0 to 279 mg (mass) for overlapping 3-mm sections (C). Calcium scores (both Agatston and mass) were nearly identical for MPR (B) and raw data approaches (A), with inter-quartile ranges of 0–1% for inter-scan variability. Median inter-scan variability with contiguous 3-mm sections (B) was 13% (Agatston) and 11% (mass). Median variability was reduced to 10% (Agatston and mass) with contiguous 1.5-mm sections (D) and to 8% (Agatston) and 7% (mass) with overlapping 3-mm MPR (A).ConclusionCalcium scoring on MPR yields nearly identical results to calcium scoring on images directly reconstructed from raw data. Overlapping MPR from prospectively ECG-triggered scans improve inter-scan variability of calcium scoring without increasing patient radiation dose.

Related Topics
Health Sciences Medicine and Dentistry Radiology and Imaging
Authors
, , ,