Article ID Journal Published Year Pages File Type
4281098 The American Journal of Surgery 2008 7 Pages PDF
Abstract

BackgroundThis randomized, double-arm trial was designed to study the benefit of a novel device (MarginProbe, Dune Medical Devices, Caesarea, Israel) in intraoperative margin assessment for breast-conserving surgery (BCS) and the associated reduction in reoperations.MethodsIn the device group, the probe was applied to the lumpectomy specimen and additional tissue was excised according to device readings. Study arms were compared by reoperation rates and by correct surgical reaction confirmed by histology.ResultsThree hundred patients were enrolled. Device use was associated with improved correct surgical reaction, defined as additional re-excision in all histologically detected positive margins, with tumor within 1 mm of inked margin. The repeat lumpectomy rate was significantly reduced by 56% in the device arm: 5.6% versus 12.7% in the control arm. There were no differences in excised tissue volume or cosmetic outcome.ConclusionsIntraoperative use of the MarginProbe for positive margin detection is safe and effective in BCS and decreases the rate of repeat operations.

Related Topics
Health Sciences Medicine and Dentistry Surgery
Authors
, , , , , , , , , , , , , , , , ,