Article ID Journal Published Year Pages File Type
4281695 The American Journal of Surgery 2006 6 Pages PDF
Abstract

BackgroundOur aim was to determine whether suprarenal fixation in endografts compromises renal artery (RA) flow and whether subsequent RA intervention is precluded by the stent struts.MethodsProspectively acquired data from 104 patients with endovascular aneurysm repair (EVAR) of an abdominal aortic aneurysm using the Zenith endograft (Cook, Inc., Bloomington, IN) were analyzed. The Zenith device uses a 26-mm, uncovered, barbed Z stent for suprarenal function.ResultsNo RA stenosis, occlusion, or infarction resulted from the suprarenal stent. In 3 of 104 (2.9%) patients, RA compromise (2 stenoses, 1 occlusion) was caused by impingement of graft material on the lowermost RA. The 2 RA stenoses were stented successfully at 1 and 7 months post-EVAR. Six of 104 (5.8%) patients developed late stenoses unrelated to the endograft: all were stented successfully from 19 to 36 months after EVAR. One patient with severe RA stenosis had balloon angioplasty pre-EVAR and then was stented electively 6 weeks post-EVAR.ConclusionsOur data show that the suprarenal fixation of the Zenith aortic endograft does not cause RA stenosis, occlusion, or infarction, nor does it preclude post-EVAR renal artery intervention.

Keywords
Related Topics
Health Sciences Medicine and Dentistry Surgery
Authors
, , , , , , , , ,