Article ID Journal Published Year Pages File Type
4337136 Journal of Neuroscience Methods 2006 9 Pages PDF
Abstract

MPScope is a software suite to control and analyze data from custom-built multiphoton laser scanning fluorescence microscopes. The acquisition program MPScan acquires, displays and stores movies, linescans, image stacks or arbitrary regions from up to four imaging channels and up to two analog inputs, while plotting the intensity of regions of interest in real-time. Bidirectional linescans allow 256 × 256 pixel frames to be acquired at up to 10 fps with typical galvanometric scanners. A fast stack mode combines movie acquisition with continuous z-focus motion and adjustment of laser intensity for constant image brightness. Fast stacks can be automated by custom programs running in an integrated scripting environment, allowing a 1 mm3 cortical volume to be sampled in 1 billion voxels in approximately 1 h. The analysis program MPView allows viewing of stored frames, projections, automatic detection of cells and plotting of their average intensity across frames, direct frame transfer to Matlab, AVI movie creation and file export to ImageJ. The combination of optimized code, multithreading and COM (Common Object Model) technologies enables MPScope to fully take advantage of custom-built two-photon microscopes and to simplify their realization.

Related Topics
Life Sciences Neuroscience Neuroscience (General)
Authors
, , ,