Abstract

•Cadaver middle ears were imaged in real-time using high frequency ultrasound.•Ossicles and their support structures were readily identifiable with 30–60 μm resolution.•Vibrations of middle ear structures were measured in response to sound stimuli.•The procedure has great potential to aid in diagnosis and monitoring of pathology.

Imaging techniques currently used in the clinic to inspect ears in patients are generally limited to views terminating at the tympanic membrane (TM) surface. For imaging past the TM, methods such as computed tomography are typically used, but in addition to disadvantages such as being costly, time consuming, and causing radiation exposure, these often do not provide sufficient resolution of the middle ear structures of interest. This study presents an investigation into the capability of high frequency ultrasound to image the middle ear with high resolution in real-time, as well as measure vibrations of TM and middle ear structures in response to sound stimuli. In unfixed cadaver ears, the TM, ossicles, and ossicular support tissues were all readily identifiable, with capabilities demonstrated for real-time imaging and video capture, and vibrometry of middle ear structures. Based on these results, we conclude that high frequency ultrasonography is a relatively simple and minimally invasive technology with great potential to provide clinicians with new tools for diagnosing and monitoring middle ear pathologies.