Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
4356345 | Hearing Research | 2006 | 12 Pages |
Abstract
The results confirm previous reports that delaying charge recovery, in this case by switching from a BP to an ALT-m wave shape, can substantially reduce thresholds [Van Wieringen, A., Carlyon, R.P., Laneau, J., Wouters, J., 2005. Effects of waveform shape on human sensitivity to electrical stimulation of the inner ear. Hear. Res. 200, 73-86; Carlyon, R.P., van Wieringen, A., Deeks, J.M., Long, C.J., Lyzenga, J, Wouters, J., 2005. Effect of inter-phase gap on the sensitivity of cochlear implant users to electrical stimulation. Hear. Res. 205, 210-224]. However, at high pulse rates, this advantage only occurs at short phase durations. In addition, we show that the complex interaction between the effects of pulse shape, rate, and phase duration on thresholds can be captured by the simple linear model described by Carlyon et al.
Keywords
Related Topics
Life Sciences
Neuroscience
Sensory Systems
Authors
Astrid van Wieringen, Robert P. Carlyon, Olivier Macherey, Jan Wouters,