Abstract

An inexpensive sensitive gas-phase chemiluminescence (GPCL) based analyzer for arsenic is described; this device utilizes manual fluid dispensing operations to reduce size, weight and cost. The analyzer in its present form has a limit of detection (LOD, S/N = 3) of 1.0 μg/L total inorganic As (peak heightbased, 3 mL sample). The system was used to measure low level arsenic in tap water samples from Texas and New Mexico and compared with results obtained by inductively coupled plasma-mass spectrometry (ICP-MS) as well as those from an automated GPCL analyzer. Good correlations were observed. Higher levels of As (50–500 μg/L, As(III), As(V) and mixtures thereof) were spiked into local tap water; the recoveries ranged from 95 ± 2% to 101 ± 1%. A single instrument weighs less than 3 kg, consumes <25 W in power, can be incorporated in a briefcase and constructed for <$US $1000. It is easily usable in the field.