Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
4455962 | Journal of Environmental Sciences | 2010 | 5 Pages |
A reliable and sensitive competitive real-time fluorescent quantitative immuno-PCR (RTFQ-IPCR) assay using a molecular beacon was developed for the determination of trace fluoranthene (FL) in the environment. Under optimized assay conditions, FL can be determined in the concentration range from 1 fg/mL to 100 ng/mL, with y = 0.194x + 7.859, and a correlation coefficient of 0.967 was identified, with a detection limit of 0.6 fg/mL. Environmental water samples were successfully analyzed, recovery was between 90% and 116%, with intra-day relative standard deviation (RSD) of 6.7%–12.8% and inter-day RSD of 8.4%–15.2%. The results obtained from RTFQ-IPCR were confirmed by ELISA, showing good accuracy and suitability to analyze FL in field samples. As a highly sensitive method, the molecular beacon-based RTFQ-IPCR is acceptable and promising for providing reliable test results to make environmental decisions.