4699398 Chemical Geology 2012 6 Pages PDF
Abstract

We present new pressure calibration experiments demonstrating that the piston cylinder apparatus is suitable for experiments at pressure as low as 150 MPa. Two newly designed 25 mm furnace assembly have been developed and calibrated using two different calibration methods: the NaCl melting curve and the solubility of H2O in albitic and rhyolitic melts. The NaCl calibration experiments performed in the pressure range 150–500 MPa yield the pressure correction that has to be applied to the nominal pressure in order to have the equivalent (real) pressure on the sample. This correction varies as a function of the experimental pressure as follows:Pcorrection(MPa)=−0.115xPnominal(MPa)+78.23Pcorrection(MPa)=−0.115xPnominal(MPa)+78.23The H2O solubility experiments in albitic and rhyolitic melts confirm the corrections determined using the NaCl calibration method. Moreover, because these experiments are performed at temperatures higher than those used for NaCl calibration, they demonstrate that the pressure correction is not affected by temperature in the range 800–1000 °C. The accuracy of the pressure estimate associated with the calibration methods is ± 25 MPa.The major advantage of using the new assemblies is that low pressure experiments, which require rapid heating and quenching rates (e.g. volcanic and hydrothermal systems), can be performed with the same ease and precision as standard high pressure experiments for which piston cylinder is routinely used.

► We have calibrated a piston cylinder for low pressure experiments (P ≥ 150 MPa). ► NaCl calibration technique allows accurate determination of pressure correction. ► Pressure correction does not vary in the temperature range 800–1000 °C.

Related Topics
Physical Sciences and Engineering Earth and Planetary Sciences Geochemistry and Petrology
Authors
, , , , , , ,