The FIS first oxidation peak showed an electrode process controlled by diffusion at 2-MES modified gold electrodes, while the electrode process showed a diffusion/adsorption mixed control at 4-MP modified gold electrodes. Square wave voltammetry was used to perform the FIS quantitative determination at both modified electrodes using the commercial reagent. Calibration curves were linear in the concentration range from 1 Ã 10â 7 to 1 Ã 10â 4 M and from 1 Ã 10â 6 to 1 Ã 10â 5 M for 2-MES and 4-MP SAM modified electrodes, respectively. Limits of detection for a 3:1 signal to noise ratio were 5 Ã 10â 8 M (14.3 ppb) and 5 Ã 10â 7 M (143 ppb) for 2-MES and 4-MP modified electrodes, respectively. The reproducibility and the repeatability were 2.2% and 0.5% for 2-MES SAM modified electrodes, and 1.6% and 1.3% for 4-MP SAM modified electrodes, respectively.