Article ID Journal Published Year Pages File Type
4985808 Tribology International 2017 29 Pages PDF
Abstract
The grasping action in a minimally invasive surgery operation is a blind operation in terms of the touch feedback and may result an undesirable tissue damage during the surgery. In this study, the interaction of the minimally invasive surgery tool with the tissue investigated by integrating a high resolution tactile sensor to the end-effector of the tool. The pressure distribution during the closing action of the grasper is characterized using the non-organic viscoelastic materials as well as organic tissues such as dead chicken liver and meat. Pressure mapping is used to calculate the net grasping force acting on the contact interface. The results show the total grasping force changes in the range of 0.2-0.7 N for different types of tissues.
Related Topics
Physical Sciences and Engineering Chemical Engineering Colloid and Surface Chemistry