Abstract
Slush nitrogen, a cryogenic fluid with solid particles suspended in liquid, is considered a potential coolant for high temperature superconductors. In this study, the capacitance-type meters, which are adopted for measuring the density, liquid level and flow velocity of slush nitrogen, are improved with their accuracy and reliability by optimizing the measurement components and mechanical structure. For density measurement, a capacitance-to-digital converter with high accuracy and resolution is used. The liquid level meter is equipped with a coaxial shielding tube, which is earth grounded to protect the electrodes from mechanical disturbance and electromagnetic interference, to strengthen the mechanical support and to improve the stability. The capacitors with double-circular curved-plate electrodes are adopted in the flowmeter to improve its sensitivity. According to the calibration results, the densimeter and the liquid level meter have high sensitivity and good linearity, with the improved accuracy of ±0.16% and ±1.0%, respectively. Besides, the flow velocity can be calculated from the capacitors' mounting spacing and the delay time of density fluctuation analyzed by cross-correlation function, which turns out to have a relative error within ±7.5% compared with that calculated by the measured variation of liquid level.
Keywords
Related Topics
Physical Sciences and Engineering
Chemical Engineering
Fluid Flow and Transfer Processes
Authors
Yijian Li, Shuqin Wu, Tao Jin, Gang Lei,