Article ID Journal Published Year Pages File Type
5006382 Measurement 2017 7 Pages PDF
Abstract
Gas flow was measured by using constant volume-variable pressure method. The gas flow is measured by pressure rise method using spinning rotor gauge (SRG). The design is relatively simple and has no precisely movable parts compared to the constant pressure-variable volume method. Due to outgassing effect we used a non-evaporable getter pump in order to diminish the influence of hydrogen outgassing on measured gas flow. Lower limit of pressure rise method is 5 × 10−10 mbar l/s and this is extended to 3 × 10−12 mbar l/s by using constant conductance method. With this flow meter we generated a calibration pressure of Argon in our calibration system in the range from 8 × 10−13 mbar up to 2 × 10−7 mbar.
Related Topics
Physical Sciences and Engineering Engineering Control and Systems Engineering
Authors
, ,