Abstract
At National Institute of Metrological Research (INRIM), a selectable-value transportable high dc voltage standard (THVS) operating in the range from 10Â V to 100Â V in steps of 10Â V, was developed. This standard was built to cover the lack of high level dc voltage standards at values higher than 10Â V to employ mainly as transportable standard for inter-laboratory comparisons (ILCs) or eventually as laboratory (local) high level dc voltage standard. An electronic technique was used to improve the THVS accuracy of the high values. The THVS shows lower noise, better short- term stability and accuracy than top level dc voltage and multifunction calibrators (MFCs) and satisfactory suitability to be transported. The project is extensible to 1000Â V.
Related Topics
Physical Sciences and Engineering
Engineering
Control and Systems Engineering
Authors
Flavio Galliana, Roberto Cerri, Luca Roncaglione Tet,