Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5032655 | Medical Engineering & Physics | 2017 | 4 Pages |
Abstract
The results showed that the proposed design ensured a high repeatability in repositioning the pots with standard deviations for the computed distance and angle between the pots at both ends of the joint equal to 0.1 mm, 0.01 mm, 0.13° and 0.03° for the tibial and femoral fixtures respectively. Therefore, it is possible to remove and re-setup a joint with high precision. The results also showed that the errors in repositioning the robotic platform (that is: specimen path repeatability) were 0.11 mm and 0.12°, respectively.