Article ID Journal Published Year Pages File Type
5032743 Medical Engineering & Physics 2017 5 Pages PDF
Abstract
AO lag screws are widely used in surgical intra-articular fracture treatment for anatomical reduction and rigid fixation. Interfragmentary compressive force (IFCF) and pull-out strength (POS) are two critical parameters generated by AO lag screws during tightening, and both of these parameters could be used to estimate screw insert conditions to prevent screw stripping. The aim of this study is to evaluate the IFCF and POS of AO cancellous screws inserted into uniform synthetic cancellous bone during tightening procedures. Seven synthetic cancellous bone blocks were used for this research. Each test contained two continuous portions as follows: the rotation test portion and the pull-out test portion. IFCF and POS were captured by the pressure transducer and the sensor of the test machine. The properties of IFCF and POS based on tightening degrees were obtained in this study. The ideal balance between POS and IFCF during screw tightening exists, and the peak values of these parameters cannot be simultaneously achieved. Moreover, rotation angles of 100-150° appear to serve as the optimum balance between IFCF and POS in the present study.
Related Topics
Physical Sciences and Engineering Engineering Biomedical Engineering
Authors
, , , , , , ,