Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5131783 | Analytical Biochemistry | 2016 | 4 Pages |
Abstract
Isothermal titration calorimetry (ITC) is one of the most robust label- and immobilization-free techniques used to measure protein - small molecule interactions in drug design for the simultaneous determination of the binding affinity (ÎG) and the enthalpy (ÎH), both of which are important parameters for structure-thermodynamics correlations. It is important to evaluate the precision of the method and of various ITC instrument models by performing a single well-characterized reaction. The binding between carbonic anhydrase II and acetazolamide was measured by four ITC instruments - PEAQ-ITC, iTC200, VP-ITC, and MCS-ITC and the standard deviation of ÎG and ÎH was determined. Furthermore, the limit of an approach to reduce the protein concentration was studied for a high-affinity reaction (Kd = 0.3 nM), too tight to be measured by direct (non-displacement) ITC. Chemical validation of the enthalpy measurements is discussed.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Vaida LinkuvienÄ, Georg Krainer, Wen-Yih Chen, Daumantas Matulis,