Abstract

•A simple LC-IDMS/MS method for the measurement of serum urea was developed.•The LC-IDMS/MS method was demonstrated to be precise and accurate.•Good comparability with JCTLM approved GC-IDMS method was demonstrated.•First report of LC-IDMS/MS method as reference procedure in EQA for clinical labs.•Clinical laboratories' results were found to be close to the LC-IDMS/MS results.

Reference measurement procedures are essential for the accurate measurement of urea in human serum. The widely used gas chromatography-isotope dilution mass spectrometry (GC-IDMS) involves tedious derivatization processes. We developed and validated a simple liquid chromatography-isotope dilution tandem mass spectrometry (LC-IDMS/MS) method for the measurement of urea in human serum.In this method, 13C,15N2-urea was spiked into serum samples as the internal standard, and pure urea certified reference material (CRM) was used as the calibration standard to ensure the traceability of the measurements to the International System of Units (SI). The sample cleanup procedure involved a simple protein precipitation step, followed by centrifugation and filtration. The developed LC-IDMS/MS method was found to be accurate and highly reproducible. The coefficient of variation (CV) was generally below 1%. Good accuracy was demonstrated by measuring serum CRMs from two National Metrology Institutes (NMIs) with relative deviations from the certified values being close to or below 1%, and measurement uncertainties of 2.0%-2.8%. The method was used to assign the reference values of serum urea in ten cycles of an external quality assessment (EQA) program for clinical laboratories over a period of four years. The robust mean of the reported results from clinical laboratories using automated analyzers were close to the assigned values from LC-IDMS/MS method.With good accuracy and precision, as well as simple sample cleanup procedure, the developed LC-IDMS/MS method can be used as a reference measurement procedure for the measurement of urea in human serum.

Graphical abstractDownload high-res image (121KB)Download full-size image

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , , , ,