Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5135217 | Journal of Chromatography A | 2017 | 21 Pages |
Abstract
In balanced anesthesia, sevoflurane and propofol are often used in combination to achieve a better anesthetic effect. However, methods for on-line monitoring of concentrations of the two anesthetics in patients are still rare in clinical. This study proposed a non-invasive method utilizing a fast gas chromatograph combined with a surface acoustic wave sensor (Fast GC-SAW) to simultaneously on-line monitor sevoflurane and propofol in patients' exhaled gas. By using the direct resistive heating capillary column, the single detection time of Fast GC-SAW system was significantly shortened to 90s, as well as the size reduced to (40cm Ã 30cm Ã 20 cm). Besides, in the calibration of sevoflurane, Fast GC-SAW system showed a good linear correlation (R2 = 0.9925, P < 0.01) with gas chromatography-mass spectrometer (GC-MS), which ensured the reliability and accuracy of the Fast GC-SAW system. Finally, clinical experiments on patients under balanced anesthesia were conducted. The varied concentrations measured by Fast GC-SAW extraordinarily matched the clinical usages of these two anesthetics.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Hao Dong, Feng Jiang Zhang, Fu Yuan Wang, Ying Ying Wang, Jing Guo, G.M. Kanhar, Jing Chen, Jun Liu, Chen Zhou, Min Yan, Xing Chen,