Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5135791 | Journal of Chromatography A | 2017 | 19 Pages |
â¢The effects of pressure of solute partition and retention in GC.â¢Improvement of predicted retention time accuracy considering pressure contribution.â¢The relevance of considering pressure in GC and FM-GCxGC.â¢A new method for stationary phase film thickness determination.â¢The accuracy of gas velocity calculation in the two columns in FM-GCxGC.
This study aims at modelling and predicting solute retention in capillary Gas Chromatography (GC) and Flow Modulation comprehensive GC (FM-GCxGC). A new thermodynamic model, taking into account the effects of temperature and pressure, is proposed to describe the variation of the equilibrium partition constant of a solute during its elution. This retention model was challenged with the classical one, and both were applied to: (i) stationary phase film thickness indirect estimation; (ii) retention time (RT) prediction of a set of 11 model polycyclic aromatic hydrocarbons (PAHs) on the SLB-IL60 and DB-35MS columns, in temperature-programmed mode. Film thickness determination led to values about 2 times lower than those indicated by column nominal dimensions, whatever the employed model. Prediction of retention times in GC led to 0.84 and 0.26% mean errors using the classical and the extended models, respectively. Prediction in GCxGC gave 5.5 and 0.44% mean errors in 1st dimension RTs, and 7.3 and 2.2% mean errors in 2nd dimension RTs, using the classical and the extended models, respectively.