Article ID Journal Published Year Pages File Type
5136075 Journal of Chromatography B 2017 7 Pages PDF
Abstract

•A multiplex quantitative analysis of eicosanoids in blood is possible using LC-MS.•Serum eicosanoids reflect platelet activation.•LC-MS/MS assays of the serum eicosanoids are reproducible.•The measurement of serum eicosanoids may be promising for clinical applications.

Eicosanoid mediators play important roles in maintaining the physiological and pathophysiological homeostasis in the body. Their measurements, however, are rarely performed in clinical practice. In the present study, we analyzed 30 varieties of eicosanoid mediators that were detectable in human plasma and serum collected from healthy donors, using liquid chromatography-tandem mass spectrometry from the viewpoint of the clinical application of the multiplex quantitation of eicosanoid mediators. Wider variety of eicosanoid mediators were detected in serum (27 out of 30) than in plasma (14 out of 30), since the serum was thought to contain lipid mediators released from activated platelets. Larger inter-individual variations were observed in the plasma and serum eicosanoid levels. On the other hand, the concentrations of eicosanoids were not affected by the platelet count but were affected by the concentration of arachidonic acid (AA) within the reference interval (17.4-40.5 × 1010/L). When serum samples from patients with hematological disorders were analyzed, the concentrations of AA were positively correlated with the platelet count. When the patients underwent ASA therapy, a marked decrease in the concentrations of thromboxane B2 (TXB2) and 12-hydroxyl-heptadecatrienoic acid (12-HHT) was observed. Considering the availability of serum samples in clinical settings, the serum analysis of eicosanoids may be clinically useful.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , ,