Article ID Journal Published Year Pages File Type
5203017 Polymer Degradation and Stability 2009 7 Pages PDF
Abstract

A method for the determination of polymer additives like antioxidants, UV absorbers and processing stabilizers using liquid chromatography (LC) coupled with atmospheric pressure photoionization mass spectrometry (APPI-MS) is presented. Ion source parameters were optimized regarding temperatures, gas flow rates, and voltages applied. Detection limits were determined using APPI with or without dopant and were compared with electrospray ionization (ESI) and atmospheric pressure chemical ionization (APCI). Differences between APPI, ESI and APCI are pointed out and the effect of the dopant toluene and acetone is discussed. The optimized method yielded detection limits between 0.001 mg L−1 and 0.022 mg L−1 for 15 different analytes. Linear calibration plots could be obtained for all solutes over a wide concentration range showing satisfying repeatability with standard deviations of peak areas between 3.4% and 7.6%. The results indicate that the developed method can be regarded as suitable for the quantitative determination of polymer additives even at low concentration levels.