Abstract
A vertically aligned carbon nanofiber-based (VACNF) electrode platform was developed for an enzymeless hydrogen peroxide sensor. Vertical nanofibers have heights on the order of 2-3 μm, and diameters that vary from 50 to 100 nm as seen by atomic force microscopy. The VACNF was grown as individual, vertically, and freestanding structures using plasma-enhanced chemical vapor deposition. The electrochemical sensor, for the hydrogen peroxide measurement in solution, showed stability and reproducibility in five consecutive calibration curves with different hydrogen peroxide concentrations over a period of 3 days. The detection limit was 66 μM. The sensitivity for hydrogen peroxide electrochemical detection was 0.0906 mA cmâ2 mMâ1, respectively. The sensor was also used for the measurement of hydrogen peroxide as the by-product of the reaction of cholesterol with cholesterol oxidase as a biosensor application. The sensor exhibits linear behavior in the range of 50 μM-1 mM in cholesterol concentrations. The surface analysis and electrochemistry characterization is presented.
Keywords
Related Topics
Physical Sciences and Engineering
Chemistry
Physical and Theoretical Chemistry
Authors
D. Suazo-Dávila, J. Rivera-Meléndez, J. Koehne, M. Meyyappan, C.R. Cabrera,