Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5361083 | Applied Surface Science | 2008 | 6 Pages |
Abstract
A comparison between roughness data obtained with an atomic force microscope (AFM) on different surfaces requires reliable roughness parameters. In order to specify the appropriate parameters for nanoscale roughness measurements, we compared the root mean square (rms) roughness and the relative surface area (sdr) as function of varying scan size, speed and pixel size. By using oxygen plasma (24 kJ) treated SU-8 with an average rms roughness of 2.6 ± 0.5 nm as reference surface, the repeatability of the method was evaluated for dynamic (tapping) and contact mode. The evaluation of AFM images indicated a decrease of the effective tip radius after a few measurements. This degradation of the tip lowers the resolution of the image and can affect roughness measurements.
Keywords
Related Topics
Physical Sciences and Engineering
Chemistry
Physical and Theoretical Chemistry
Authors
Ferdinand Walther, Wolfgang M. Heckl, Robert W. Stark,