Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5395681 | Journal of Electron Spectroscopy and Related Phenomena | 2015 | 9 Pages |
Abstract
A new type of electrostatic electron spectrometer is developed, capable of particular sensitive measurements of energy spectra and time-of-flight distributions. This instrument is specifically designed and optimized for laser-pump/X-ray-probe measurements, where photo electrons or Auger electrons from surfaces, clusters, molecular or atomic targets are being measured with high time-resolution at an extremely low detection-noise level. The compact and robust cylinder-symmetrical system is a strongly improved Bessel-Box design, featuring electron retardation, a large detection solid-angle, about 100% electron transmission (gridless design) and excellent time-resolution. In this paper we describe the principle of operation of this type of spectrometer and various tests. We present quantitative results for electron measurements with different solid-state targets and two different electron-detection systems in comparison to electron-trajectory simulations inside the electrostatic spectrometer fields. Picosecond-pump/probe operation has been tested with high laser power and even the ability to work under femtosecond-pump-probe conditions with electron detection at the BESSY II slicing facility has been proven.
Keywords
Related Topics
Physical Sciences and Engineering
Chemistry
Physical and Theoretical Chemistry
Authors
G. Schiwietz, M. Beye, D. Kühn, G. Xiao,