A new method for rapid NMR imaging dubbed FLASH (fast low-angle shot) imaging is described which, for example, allows measuring times of the order of 1Â s (64Â ÃÂ 128 pixel resolution) or 6Â s (256Â ÃÂ 256 pixels). The technique takes advantage of excitation pukes with small hip angles eliminating the need of waiting periods in between successive experiments. It is based on the acquisition of the free induction decay in the form of a gradient echo generated by reversal of the read gradient. The entire imaging time is only given by the number of projections desired times the duration of slice selection and data acquisition. The method results in about a 100-fold reduction in measuring time without sacrificing spatial resolution. Further advantages are an optimized signal-to-noise ratio, the applicability of commercial gradient systems, and the deposition of extremely low rf power. FLASH imaging is demonstrated on phantoms, animals, and human extremities using a 2.3Â T 40Â cm bore magnet system. 1H NMR images are obtained with variable relaxation time contrasts and without motional artifacts.