Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5407575 | Journal of Magnetic Resonance | 2006 | 8 Pages |
Abstract
Accurate measurements of the diffusion coefficients, including an estimate of uncertainty, of various solvent molecules using the PFG-NMR method were performed in this study. Accurate diffusion coefficients were obtained using the Shigemi NMR tube. The relative combined standard uncertainties of the diffusion coefficients were found to be within approximately 0.4%. The three uncertainty sources (signal decay of the standard and the solvent, and diffusion coefficient of standard) equally affect the combined standard uncertainties. Unreliable data were obtained using a normal NMR tube, indicating that convection and background gradient effects significantly affected the accurate measurement of the diffusion coefficients.
Related Topics
Physical Sciences and Engineering
Chemistry
Physical and Theoretical Chemistry
Authors
Haruhisa Kato, Takeshi Saito, Mami Nabeshima, Kayori Shimada, Shinichi Kinugasa,