Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5407606 | Journal of Magnetic Resonance | 2006 | 8 Pages |
Abstract
The potential advantages of the higher-order resonant modes of a low-pass birdcage coil for parallel imaging are investigated using FDTD simulations in a spherical phantom. Better parallel imaging performance can be achieved in axial planes than in sagittal planes. If more modes are employed, the average g-factor (gmean) and maximum g-factor (gmax) will be improved for a specific acceleration factor (R). G-factor performance at 3 T (gmean = 1.79, gmax = 3.55) and 7 T (gmean = 1.60, gmax = 2.45) can be achieved even with an acceleration factor as high as four when all six order modes of 12-rung low-pass birdcage coil are incorporated for imaging on a mid-axial plane. For a specific number of channels, the optimum combination of corresponding modes can be obtained for different acceleration factors. Based on the g-factor and SNR performance, the total degenerate multi-mode birdcage coil with six order resonant modes has better homogeneous coverage and SENSE performance than the 8-element phased array coil, although requiring fewer channels. In addition, the dielectric effects at high field can improve the parallel imaging performance.
Related Topics
Physical Sciences and Engineering
Chemistry
Physical and Theoretical Chemistry
Authors
Chunsheng Wang, Peng Qu, Gary X. Shen,