Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5434371 | Materials Science and Engineering: C | 2017 | 6 Pages |
â¢In vivo degradation of Mg-Nd-Zn-Zr (JDBM) DESs was reported in porcine coronary arteries in this work.â¢The JDBM DESs degraded slowly and kept most of the radial supporting strength post implantation for 3 monthsâ¢The JDBM DESs showed favourable biosafety, biocompatibility and neointimal proliferation.
In this work, rapamycin-eluting poly (d, l-lactic acid) coating (PDLLA/RAPA) was prepared on biodegradable Mg-Nd-Zn-Zr alloy (JDBM) for both in vitro and in vivo investigation of the degradation behaviors of the magnesium alloy stent platform. Electrochemical tests and hydrogen evolution test demonstrated significant in vitro protection of the polymeric coating against magnesium degradation both in short and long term. The 3-month in vivo study on the RAPA-eluting JDBM stent implanted into porcine coronary arteries confirmed its favorable safety, and in the meanwhile revealed similar neointima proliferation compared to the second generation DES Firebird 2 with no occurrence of adverse complications. Moreover, Micro-CT examination combined with IVUS and OCT detection indicated a remarkably lower degradation rate and prolonged radial supporting duration of the drug-eluting JDBM stent as compared to the bare, attributable to the protection of the coating in vivo. Hence, rapamycin-eluting JDBM stents exhibit great potential for clinical application.
Graphical abstractDownload high-res image (130KB)Download full-size image