Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5449293 | Optics Communications | 2017 | 5 Pages |
Abstract
A microfluidic-based multi-angle laser scattering (MALS) system capable of acquiring scattering patterns of a single particle is designed and demonstrated. The system includes a sheathless nozzle microfluidic glass chip, and an on-chip MALS unit being in alignment with the nozzle exit in the chip. The size and relative refractive indices (RI) of polystyrene (PS) microspheres were deduced with accuracies of 60  nm and 0.002 by comparing the experimental scattering patterns with theoretical ones. We measured scattering patterns of waterborne parasites i.e., Cryptosporidium parvum (C.parvum) and Giardia lamblia (G. lamblia), and some other representative species suspended in deionized water at a maximum flow rate of 12  μL/min, and a maximum of 3000 waterborne parasites can be identified within one minute with a mean accuracy higher than 96% by classification of distinctive scattering patterns using a support-vector-machine (SVM) algorithm. The system provides a promising tool for label-free detection of waterborne parasites and other biological contaminants.
Keywords
Related Topics
Physical Sciences and Engineering
Materials Science
Electronic, Optical and Magnetic Materials
Authors
Wei Huang, Limei Yang, Lei Lei, Feng Li,