Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5449447 | Optics Communications | 2017 | 4 Pages |
Abstract
Spectrometers are generally sensitive to the polarization state of the measured light, which affects their measurement accuracy. In this study, we developed a method to reduce the polarization sensitivity by using an achromatic quarter-wave plate in a double-pass optical structure. By rotating the polarization state of the second pass beam with an achromatic quarter-wave plate, the diffraction efficiencies of the grating for the s-polarized and p-polarized light become nearly the same. The formulas of this polarization-insensitive system were strictly derived. We designed a double-pass spectrometer with a grating of 1050 lines/mm and a reflective achromatic quarter-wave plate. The results of the numerical simulation show that the polarization-dependent loss of the designed spectrometer is less than 0.5Â dB.
Related Topics
Physical Sciences and Engineering
Materials Science
Electronic, Optical and Magnetic Materials
Authors
Yajun Pang, Yinxin Zhang, Huaidong Yang, Zhanhua Huang,