We present the optical design of a Czerny-Turner imaging spectrometer for which astigmatism is corrected using off-the-shelf optics resulting in spectral resolution of 0.1 nm. The classic Czerny-Turner imaging spectrometer, consisting of a plane grating, two spherical mirrors, and a sensor with 10-μm pixels, was used as the benchmark. We comparatively assessed three configurations of the spectrometer that corrected astigmatism with divergent illumination of the grating, by adding a cylindrical lens, or by adding a cylindrical mirror. When configured with the added cylindrical lens, the imaging spectrometer with a point field of view (FOV) and a linear sensor achieved diffraction-limited performance over a broadband width of 400 nm centered at 800 nm, while the maximum allowable bandwidth was only 200 nm for the other two configurations. When configured with the added cylindrical mirror, the imaging spectrometer with a one-dimensional field of view (1D FOV) and an area sensor showed its superiority on imaging quality, spectral nonlinearity, as well as keystone over 100 nm bandwidth and 10 mm spatial extent along the entrance slit.