Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5452900 | Clinical Plasma Medicine | 2017 | 34 Pages |
Abstract
In summary, the results of the experiments indicate a high potential of the plasma jet to be used as a medical device exhibiting low gas temperatures up to 34 °C. The calculated leakage currents are mostly below the 10 µA threshold. The limiting UV exposure duration for the APJ with a calculated maximum effective irradiance of 2.6 µW/cm2 is around 19 min, based on the exposure limits of the international commission on non-ionizing radiation protection guidelines (ICNIRP) [2]. A significant ozone concentration was observed mainly in the axial effluent gas flow. Ozone concentration strongly decreases with increasing distance from the plasma source exit nozzle.
Keywords
Related Topics
Physical Sciences and Engineering
Materials Science
Materials Science (General)
Authors
A. Lehmann, F. Pietag, Th. Arnold,