Abstract
												The dosimetric properties such as dose response, minimum detectable dose, energy response, reusability, fading properties, thermal stability and effect of reading temperatures on OSL signals were examined. OSL signals of CaSO4:Eu pellets were decreased by approximately 8% at the end of the 24 h and by about 7% at the end of 28 days when compared with the first readout. The thermal stability of the â¼240 °C TL peak and OSL signal using isothermal decay measurements were used to determine the trap parameters. The CaSO4:Eu OSL dosimeter in accordance with the presented study allows a high-sensitivity OSL phosphor whose host material was previously known as conventional synthetic CaSO4 TLD material.