Abstract
Ultrafast Ultrasound Localization Microscopy uses microbubbles that are individually localized with a resolution below 10 μm. Positions of the microbubbles are accumulated to create a super resolution image, which bypass the diffraction-limit of spatial resolution. However, microbubbles localization is affected by physiological motions at the micrometric scale. Here, we demonstrate a phase correlation method for rigid motion correction. Spatiotemporal filters extract tissue dominated images, which are tracked to correct linear motions and improve the precision of microbubbles' localization, improving the quality of the image. It is the first proof of concept towards a full motion correction strategy and super-resolution imaging in moving tissues.