Article ID Journal Published Year Pages File Type
5485723 Ultrasound in Medicine & Biology 2017 6 Pages PDF
Abstract
We investigated the correlation between hepatic venous pressure gradient (HVPG) and liver shear wave speed (SWS) measured by multi-frequency time-harmonic ultrasound elastography (THE) before and after transjugular intrahepatic portosystemic shunt (TIPS) implantation. Ten patients with ascites, cirrhotic liver disease and portal hypertension were prospectively examined with invasive HVPG measurement and THE before and after TIPS implantation. HVPG and SWS decreased after TIPS placement from 20.4 ± 2.2 mmHg to 9.8 ± 4.1 mmHg (mean ± standard deviation) and from 3.87 ± 0.54 m/s to 3.27 ± 0.44 m/s. Mean reduction HVPG was −10.6 ± 3.7 mmHg, p < 0.001; mean reduction SWS was −0.60 ± 0.29 m/s, p < 0.001. A linear correlation was observed between HVPG and SWS (R = 0.59, p = 0.0061). THE-measured SWS is a first potential direct ultrasound marker for liver decompression following TIPS in ascites-associated cirrhotic liver disease and therefore might be suitable to non-invasively detect portal hypertension.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Acoustics and Ultrasonics
Authors
, , , , , , ,