Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5485948 | Ultrasound in Medicine & Biology | 2017 | 8 Pages |
Abstract
The incidence of breast cancer is increasing worldwide, reinforcing the importance of breast screening. Conventional hand-held ultrasound (HHUS) for breast screening is efficient and relatively easy to perform; however, it lacks systematic recording and localization. This study investigated an electromagnetic tracking-based whole-breast ultrasound (WBUS) system to facilitate the use of HHUS for breast screening. One-hundred nine breast masses were collected, and the detection of suspicious breast lesions was compared between the WBUS system, HHUS and a commercial automated breast ultrasound (ABUS) system. The positioning error between WBUS and ABUS (1.39 ± 0.68 cm) was significantly smaller than that between HHUS and ABUS (1.62 ± 0.91 cm, p = 0.014) and HHUS and WBUS (1.63 ± 0.9 cm, p = 0.024). WBUS is a practical clinical tool for breast screening that can be used instead of the often unavailable and costly ABUS.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Acoustics and Ultrasonics
Authors
Chiun-Sheng Huang, Ya-Wen Yang, Rong-Tai Chen, Chung-Ming Lo, Chao Lo, Ching-Fen Cheng, Chao-Shuan Lee, Ruey-Feng Chang,