Abstract
The end-cap time-of-flight (ETOF) system for the Beijing Spectrometer III (BESIII) has been upgraded using the Multi-gap Resistive Plate Chamber (MRPC) technology (Williams et al., 1999; Li et al., 2001; Blanco et al., 2003; Fonte et al., 2013, [1-4]). A set of quality-assurance procedures has been developed to guarantee the performances of the 72 mass-produced MRPC modules installed. The cosmic ray batch testing show that the average detection efficiency of the MRPC modules is about 95%. Two different calibration methods indicate that MRPCs' time resolution can reach 60 ps in the cosmic ray test.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
Z. Liu, X. Li, Y.J. Sun, C. Li, Y.K. Heng, T.X. Chen, H.L. Dai, M. Shao, S.S. Sun, Z.B. Tang, R.X. Yang, Z. Wu, X.Z. Wang,