Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5492665 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2017 | 8 Pages |
Abstract
We have developed a charge-to-time converting (QTC) leading-edge discriminator for plastic-scintillator signals, which is packaged in a NIM-standard double-width module. The pulse width of the discriminator output is approximately proportional to the charge integration of input signals up to 40-50 pC. The non-linearity is 20%, 10%, 4%, and 2% at maximum for the input charges below 7 pC, for 7-10 pC, for 10-20 pC, and above 20 pC, respectively, for the minimum input charges larger than 2 pC. The discriminator could be suitable in the measurement of the timing and charge of plastic-scintillator signals together with a multi-hit time-to-digital converter (TDC) module.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
T. Ishikawa, Y. Takeda, Y. Honda, Y. Inoue, H. Kanda, S. Kido, Y. Matsumura, M. Miyabe, I. Nagasawa, H. Shimizu, T. Takeda, A.O. Tokiyasu, Y. Tsuchikawa, H. Yamazaki,