Article ID Journal Published Year Pages File Type
5492725 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2017 7 Pages PDF
Abstract
In this study, we present a symmetric charge division resistive network and associated software methods to assess the performance of a PET detector. Our approach consists in keeping the n lines and n columns information provided by a symmetric charge division circuit (SCD). We provided equations relative to output currents of the network, which enable estimation of the charge. We propose a novel approach to reconstruct the charge distribution from the lines and columns projection using a maximum likelihood expectation maximization (MLEM) approach which takes the non-uniformity of the photodetector channel gains into account. We also introduce a mathematical proof of the relation between the sigma of the reconstructed charge distribution and the Ratio between the line of interest (maximum value) and the background signal charges. To the best of our knowledge, this is the first study reporting these equations. Preliminary results obtained with a resistive network used in readout of a monolithic 50×50×8mm3 LYSO crystal coupled to a H9500 PMT validated the effectiveness of the reconstructed charge distribution to optimize both the x and y spatial resolution and the energy resolution. We obtained a mean x and y spatial resolution of 1.10 mm FWHM and a 14.7% energy resolution by calculating the integral of the reconstructed charge distribution. Finally, the relation between the ratio and the sigma of the reconstructed charge distribution may provide new opportunities in terms of Depth-of-Interaction estimation when using a monolithic crystal coupled to a multi-channel photodetector.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, , , ,