| Article ID | Journal | Published Year | Pages | File Type | 
|---|---|---|---|---|
| 5492766 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2017 | 10 Pages | 
Abstract
												We report the characterization, calibration and performance of a custom-built hybrid detector consisting of BC501A liquid scintillator and BC702 scintillator for the detection of fast and thermal neutrons, respectively. Pulse Shape Discrimination techniques are developed to distinguish events due to γ-rays, fast and thermal neutrons. Software analysis packages are developed to derive raw neutron energy spectra from measured proton recoil spectra. The validity is demonstrated through the reconstruction of the 241AmBe(α,n) neutron spectrum.
Related Topics
												
													Physical Sciences and Engineering
													Physics and Astronomy
													Instrumentation
												
											Authors
												M.K. Singh, A. Sonay, M. Deniz, M. AÄartıoÄlu, G. Asryan, G. Kiran Kumar, H.B. Li, J. Li, F.K. Lin, S.T. Lin, V. Sharma, L. Singh, V. Singh, V.S. Subrahmanyam, A.K. Soma, H.T. Wong, S.W. Yang, I.O. Yıldırım, M. Zeyrek, 
											