Article ID Journal Published Year Pages File Type
5493095 Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2017 4 Pages PDF
Abstract
Measurement of the Time-of-Flight (TOF) of the 511 keV gammas brings an important reduction of statistical noise in the PET image, with higher precision time measurements producing clearer images. The common method of coupling a photodetector to scintillating crystals is to have two matching matrices, with a one-to-one coupling between the crystal and the photodetector. We propose a new geometry based on analogue strip SiPMs reading out a scintillator cut into slabs. This technique allows the time stamping of individual photo-electrons and extracts the best time resolution using a specific algorithm. Here we present the results from the first 'slab module' test.
Related Topics
Physical Sciences and Engineering Physics and Astronomy Instrumentation
Authors
, ,