Abstract
The feasibility of the measurement instrument was tested both in laboratory with a signal generator and in the Minerve reactor. It is shown that the instrument provides reliable and robust count rate estimation over a wide reactor power range based on the third order statistics of the fission chamber signal. In addition, the system is able to identify whether the measured count rate change is due to the malfunction of the detector or due to the change in the neutron flux. The applied self-monitoring method is based on the spectral properties of the fission chamber signal. During the experimental verification, the considered malfunction was the change of the polarization voltage.