Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5493396 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2017 | 16 Pages |
Abstract
Intraoperative electron radiation therapy (IOERT), which uses electron beams for irradiating the target directly during the surgery, has the advantage of delivering a homogeneous dose to a controlled layer of tissue. Since the dose falls off quickly below the target thickness, the underlying normal tissues are spared. In selecting the appropriate electron energy, the accuracy of the target tissue thickness measurement is critical. In contrast to other procedures applied in IOERT, the routine measurement method is considered to be completely traditional and approximate. In this work, a novel mechanism is proposed for measuring the target tissue thickness with an acceptable level of accuracy. An electronic system has been designed and manufactured with the capability of measuring the tissue thickness based on the recorded electron density under the target. The results indicated the possibility of thickness measurement with a maximum error of 2Â mm for 91.35% of data. Aside from system limitation in estimating the thickness of 5Â mm phantom, for 88.94% of data, maximum error is 1Â mm.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
Mohammad Reza Yazdani, Saeed Setayeshi, Hossein Arabalibeik, Mohammad Esmaeil Akbari,