Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5493452 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2016 | 10 Pages |
Abstract
A dual-phase xenon time-projection chamber was built at Nikhef in Amsterdam as a direct dark matter detection R&D facility. In this paper, the setup is presented and the first results from a calibration with a 22Na gamma-ray source are presented. The results show an average light yield of (5.6±0.3) photoelectrons/keV (calculated to 122 keV and zero field) and an electron lifetime of (429±26) μs. The best energy resolution ÏE/E is (5.8±0.2)% at an energy of 511 keV. This was achieved using a combination of the scintillation and the ionization signals. A photomultiplier tube gain calibration technique, based on the electroluminescence signals occurring from isolated electrons, is presented and its advantages and limitations are discussed.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
E. Hogenbirk, J. Aalbers, M. Bader, P.A. Breur, A. Brown, M.P. Decowski, C. Tunnell, R. Walet, A.P. Colijn,