Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
5493454 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment | 2016 | 21 Pages |
Abstract
Five SiC Schottky photodiodes for X-ray detection have been electrically characterized at room temperature. One representative diode was also electrically characterized over the temperature range 20°C to 140 °C. The performance at 30 °C of all five X-ray detectors, in both current mode and for photon counting X-ray spectroscopy was investigated. The diodes were fabricated in an array form such that they could be operated as either a 2Ã2 or 1Ã3 pixel array. Although the devices showed double barrier heights, high ideality factors and higher than expected leakage current at room temperature (12 nA/cm2 at an internal electric field of 105 kV/cm), they operated as spectroscopic photon counting soft X-ray detectors uncooled at 30 °C. The measured energy resolution (FWHM at 17.4 keV, Mo Kα) varied from 1.36 to 1.68 keV among different diodes.
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Instrumentation
Authors
G. Lioliou, H.K. Chan, T. Gohil, K.V. Vassilevski, N.G. Wright, A.B. Horsfall, A.M. Barnett,