Abstract
In the frame of the European metrological research project MetroERM, a compact portable aerosol sampling and measurement device was developed at Jožef Stefan Institute. The system incorporates a CeBr3 scintillation detector positioned centrally within a concertinaed filter assembly and an improved high flow rate air pump. It provides continuous on-line low level airborne radioactive particulate monitoring for field station use via 3 G network communications. The calibration of the device was performed at National Physical Laboratory (NPL) with filters, spiked with a certified mixed nuclide solution. Additionally first tests were performed in an environment with an elevated radon concentration.
Keywords
Related Topics
Physical Sciences and Engineering
Physics and Astronomy
Radiation
Authors
Denis GlaviÄ-Cindro, Drago Brodnik, Toni PetroviÄ, Matjaž Vencelj, DuÅ¡an Ponikvar, Steven James Bell, Lynsey Keightley, Selina Woods,