Abstract
In this work, verification data for 11 HPGe detectors from two different manufacturers and three different intrinsic configurations were analyzed in respect to the stability of both the efficiency and resolution for the 122Â keV peak from 57Co and the 1332Â keV peak from 60Co. The results allow a discussion about the stability of these parameters over time (in some cases, almost 15 years), their sensitivity to imminent detector failures and their performance after a failure has been corrected; moreover, the results show a clear correlation between the manufacturer or configuration and the long-term performance of the detector.